2459-
An instrument is busy. A delay in restarting this application could result in loss
of new data. Continue with program Exit?
Cause:
Action:
2460-
Fatal Communications error on (Unit n - S/N: nnnn)
Cause:
Action:
2461-
No instruments are in operation. This application will unconditionally
terminate.
Cause:
Action A:
Action B:
539-42822-01 (Rev A) - Nov 2017
You attempted to exit the Elzone analysis program while an analysis is
in progress. While this is possible, the data collected while the analysis
program is inactive will not be permanently recorded until the analysis
program is re-started. A power failure to the instrument could cause
some data to be lost.
If your are not concerned with potential loss of data if a power failure
occurs, click
Yes
to continue with the exit; click
for the analysis to finish.
There was a fatal error in the communications between the analysis
program and the analyzer. All displays for the analyzer will close.
Ensure that the analyzer's ethernet connections are properly connected.
Stop and restart the analysis program. If the restarted application can-
not establish communications with the analyzer, turn the analyzer off,
then back on. If this error persists, contact your Micromeritics service
representative.
An analyzer must be active for the application to operate. The initial-
ization of the analyzes configured with the Setup program has failed.
The application stops.
Usually this message is preceded by another message giving the reason
for the analyzer's failure to initialize. See the instructions for that mes-
sage.
Ensure that the analyzer is attached to the computer. Verify that the
analyzer has the power switch in the ON position. If the application
continues to fail in its attempts to initialize the analyzer, contact your
service representative.