Elzone II 5390 Confirm
6180-
Cannot read the background file (file name).
Cause:
Action A:
Action B:
Action C:
6181-
The background file (file name) was not analyzed with the same liquid as this
file.
Cause:
Action:
6182-
There is no full-resolution data in the background file (file name).
Cause:
Action:
6183-
The orifice ID for (name) does not match the current tube's ID (name).
6184-
The orifice diameter for (file name) (number) does not match the current tube's
diameter (number).
Cause:
Action:
A-42
The selected background file is not present or cannot be opened.
Ensure that the path name is correct.
Check directory permissions.
Make sure you have sufficient space on your disk.
You selected a background that has a different liquid from the one in
the current file.
Choose a background that was performed used the same electrolyte as
the current file.
You selected a background that does not contain full-resolution data.
Choose a different background file. Do not delete full-resolution data
from background files.
You selected Calibration copied from sample (file name) and you
chose a file for an orifice tube other than the one you are using for the
current analysis.
Choose a file that contains calibration information for the current tube.