Micromeritics Elzone II 5390 Operator's Manual page 56

Advertisement

Elzone II 5390 Confirm
12. Close the sample compartment door on the right side; adjust the light and camera if
needed.
13. Select Unit > Fill System; the Fill System dialog is displayed.
This step fills the internal glassware.
14. Click
Start
closes automatically when the glassware is filled.
15. Add the number of drops specified for each reference material (refer to the booklet)
to the sample beaker.
16. Close the sample compartment door on the front side.
3-20
; a dialog indicating the operation is in progress is displayed. The dialog
Calibrating an Orifice Tube
539-42822-01 (Rev B) - Nov 2017