Appendix A
6214-
The noise level or curvature in the blending range is too high.
Cause:
Action:
6230-
The reference material ID (name), Lot (number) is already in the list.
Cause:
Action:
6231-
Please select at least one reference material.
Cause:
Action:
6232-
No more than ten reference materials can be selected.
Cause:
Action:
6233-
The reference material is outside valid range for selected orifice tube. Do you
want to add the reference material to global list only?
Cause:
Action:
539-42822-01 (Rev A) - Nov 2017
The data in the blending range are not comparable due to excessive
scatter or high curvature for at least one pair of data sets.
Collect more data or choose an additional data set to cover the blending
range.
You attempted to add a reference material that already exists in the list.
Change the identification of the material you are adding so that it does
not conflict with the one in the list.
You attempted to start an orifice tube calibration, but no reference
materials have been selected.
Select the appropriate reference material(s) and try again.
You attempted to start an orifice tube calibration with more than 10 ref-
erence materials selected.
Select the appropriate reference materials, but no more than 10.
You attempted to add a reference material that is outside of the range
for the orifice tube selected for this calibration.
Click
Yes
to add the reference material to the list. Click
reference material parameters.
Elzone II 5390 Confirm
No
to edit the
A-47