Elzone II 5390 Confirm
6051-
Flow Control: Differential pressure out of range for zero. Actual: (number)
mmHg Allowed: (number) mmHg
Cause:
Action:
6052-
Flow Control: Home detected out of range at (number) of steps, maximum
allowed (number) of steps.
Cause:
Action:
6053-
Flow Control: Waste liquid reservoir full. It must be emptied before continuing.
Cause:
Action:
6054-
Flow Control: Waste liquid full sensor not present. It must be installed before
continuing.
Cause:
Action:
6055-
Flow Control: Waste liquid reservoir overflowed. It must be emptied before
continuing.
Cause:
Action:
A-30
The pressure transducer is operating out of specification.
Contact your Micromeritics Service representative.
The volumetric pump is operating out of specification.
Contact your Micromeritics Service representative.
The waste bottle is almost full of expended electrolyte.
Empty the waste bottle and reattach it to the analyzer before beginning
another operation.
The sensor for the waste container is not properly installed.
Insert the sensor cable and ensure that it is fully seated into the con-
nector. If you continue to get this message, contact your Micromeritics
Service representatives.
Liquid has gotten into the tubing from the waste bottle to the vacuum
pump. Further operation could damage the vacuum pump.
Contact your Micromeritics Service representative.