Appendix A
6114-
You chose to use the most recent background but no background is available.
Choose Yes to proceed without background data.
Cause:
Action:
6115-
Background subtraction is selected but no background data is specified. Select
a data file from the Data Setup dialog if you wish to subtract a background.
Cause:
Action:
6116-
Selected orifice tube or liquid is different from last used on this unit. Do you
want to change the orifice tube or liquid before continuing with the analysis?.
Cause:
Action:
6117-
Please raise the beaker stand and press OK to continue, or Cancel.
Cause:
Action:
539-42822-01 (Rev A) - Nov 2017
You selected Use most recent background on the Data Setup dialog
(from the Analysis Conditions dialog) but a background has not been
performed on this instrument for the selected orifice tube and liquid.
Click
to proceed without a background. Then run a background
Yes
after the analysis. Open the sample file and click
Collected Data dialog. From the Data Setup dialog, click
choose the background file.
Click
to stop the analysis so that you can run a background, then
No
restart the analysis.
You requested background subtraction but you did not select a back-
ground file.
Click
, then click
Data Setup
select the Use most recent background option.
You selected to use an orifice tube or a liquid that is not the same as
used in the previous analysis (or calibration).
Click
to change the orifice tube or liquid. Click
Yes
with the operation; this is assuming that you have already changed the
tube or liquid.
You attempted to start an analysis with the sample stand in the lowered
position.
Raise the sample stand, then click OK to continue the analysis.
Elzone II 5390 Confirm
Data Setup
to choose a background file, or
Browse
on the
and
Browse
to continue
No
A-37