Elzone II 5390 Confirm
2477-
Unit n - S/N: nnnn did not properly initialize.
Cause A:
Action A:
Cause B:
Action B:
2478-
Error copying sequential data segment.
Cause:
Action:
2479-
Unit n - S/N: nnnn The instrument is busy performing an operation of which
this application is unaware. Do you want to cancel?
Cause:
Action:
2480-
File (name) cannot be analyzed. It is currently being edited.
Cause:
Action:
A-10
There were communication problems between the computer and the
analyzer.
Check the communication cables and make sure they are connected
securely.
You selected
Cancel
during initialization.
None required.
An internal processing and/or hardware error occurred while accessing
a portion of a sample file.
Confirm that the media being accessed does not contain errors; for
example, you may wish to use a utility such as ScanDisk.
Contact your service representative if you continue to receive this mes-
sage.
During initialization, the instrument was found to be analyzing, but the
analysis program has no record of it.
Click
to cancel the analysis and synchronize the analysis program
Yes
with the instrument's status. Click
allow the analyzer to continue the analysis.
You attempted to start an analysis using a file that is open for editing.
Finish editing the file, save and close it; then restart the analysis.
to abort the initialization and
No
539-42822-01 (Rev A) - Nov 2017