Appendix A
6185-
The calibrations could not be updated. The instrument used for the analysis
(s/n) is not in operation.
Cause:
Action:
6186-
Cannot copy calibration from this file. Orifice Tube ID (name) and diameter
(number) must match the current settings.
Cause:
Action:
6187-
Cannot open the data file (file name) for reading.
Cause:
Action A:
Action B:
6188-
Cannot create a new test in (file name).
Cause:
Action:
6189-
Cannot add a test to (file name)
Cause:
Action:
539-42822-01 (Rev A) - Nov 2017
You attempted to update calibration information but the instrument
used in the calibration is not the current instrument.
Update the calibration information using the instrument on which the
calibration was performed.
You selected a file that contains calibration parameters for an orifice
tube other than the current one.
Choose a file containing calibration information for the current orifice
tube.
You attempted to open a data file but were unsuccessful.
Check directory permissions.
Make sure the disk is not full.
An unexpected error occurred.
Contact your Micromeritics Service representative.
You have saved the maximum number of 8 tests for this file.
Delete unwanted tests, or create a new file.
Elzone II 5390 Confirm
A-43