Download Print this page

Micromeritics Elzone II 5390 Operator's Manual page 203

Advertisement

Show Status
Show Status
Select this option to show only the status window of an operation in progress.
You may wish to use this option if:
• You have an automatic operation in progress and you wish to edit a sample file, for
example. This way, you can monitor the operation while performing another task.
• You have two units attached to your computer. You can select Show Status on each
unit menu and view the status of the operations on both units at the same time.
The following items display in the status window:
Sample ID
Test
Plumbing
Data Collection
539-42822-01 (Rev A) - Nov 2017
The path and name of the sample file being used with the
current analysis.
The current test being used with the sample file; you can
perform up to eight tests with each sample file.
The current plumbing operation: Draining, Filling,
Flushing, Rinsing. Idle displays when a plumbing oper-
ation is not in progress.
The current operation in preparation for data acquisition
or calibration: Optimizing Settings, Blockage Detec-
tion, Checking Blockage, Testing Conductivity, Col-
lecting Data, Calibrating. Idle displays when an
operation is not in progress.
Elzone II 5390 Confirm
5-43