Installing the Hydropulser
5.
Fill a squeeze bottle with clean, filtered electrolyte or water and thoroughly rinse the
hydropulser tube, the orifice tube, and the electrode.
into the hydropulser tube.
Do not squeeze electrolyte up into the hydropulser tube. Liquid may get into the
hydropulser assembly and damage the assembly, or cause erratic results. Remove
the tube, clean, and dry it thoroughly if this occurs.
6.
Place your sample into the beaker and start your analysis.
Monthly
The hydropulser tube should be removed and thoroughly cleaned on a monthly basis. It is
good practice to perform this procedure on the first of each month so it will be easy to
remember.
1.
Lower the sample platform.
2.
Remove the retaining clip from the orifice tube; it is not necessary to remove it from
the hydropulser tube.
3.
Grasp the tube and remove it from its barbed fitting.
539-42822-01 (Rev A) - Nov 2017
Elzone II 5390 Confirm
Do not squeeze electrolyte up
8-17