Appendix A
6056-
Flow Control: Waste liquid overflow sensor not present. It must be installed
before continuing.
Cause:
Action:
6058-
Flow Control: Differential pressure out of range. Actual: (number) mmHg
Expected: (number) mmHg
Cause:
Action:
6059-
Flow Control: Volumetric pump home position found at (number) steps.
6060-
Flow Control: Volumetric pump home position not found at (number) steps.
6061-
Flow Control: Volumetric pump position (number) steps exceeds maximum
position (number) steps.
6062-
Flow Control: Volumetric pump movement stopped. Move was (number) steps
at (number) steps/second, direction (number), stopped after (number) seconds.
6063-
Flow Control: Volumetric pump movement timed out. Move was (number) steps
at (number) steps/second, direction (number), stopped after (number) seconds.
Cause:
Action:
6064-
Flow Control: Flow operation superseded. Old operation code: (number) New
operation code: (number)
Cause:
Action:
539-42822-01 (Rev A) - Nov 2017
The overflow sensor is not properly installed.
Contact your Micromeritics Service representative.
The differential pressure system is unable to set the required pressure.
Verify that the waste bottle top is tightly screwed onto the bottle, and
that the two vacuum lines are securely attached to the fittings on the
waste bottle top and to the fittings on the side panel of the instrument.
Verify that an orifice tube is securely attached to the orifice tube port,
and that the orifice is immersed in a beaker of electrolyte.
Restart the operation; if the message persists, contact your Micromer-
itics Service representative.
The volumetric pump has malfunctioned.
Contact your Micromeritics Service representative.
A new flow operation was started before the previous operation had
completed. The previous operation will not be completed.