Elzone II 5390 Confirm
6235-
The reference material is out of range of the orifice tube and cannot be used in
the calibration.
Cause:
Action:
6240-
The input file (file name) has too many data points. Only the first (number) lines
will be converted.
Cause:
Action:
6241-
The input file (file name) does not have enough data points. At least two size-
value pairs must be present.
Cause:
Action:
6242-
Converting data failed. Check the contents of the input file (file name).
Cause:
Action:
6243-
A negative particle size was found in the input file (file name).
Cause:
Action:
A-48
You selected a reference material that is too large or too small for the
selected orifice tube.
Select the appropriate-sized reference material(s) or orifice tube and
try again.
The number of size classes in reduced data is limited to 300; you
attempted to convert a file containing more than this number.
Reduce the number of data points in the file.
Reduced data must have at least four size classes; you attempted to
convert a file containing fewer than this number.
Increase the number of data points in the file.
You attempted to convert an ASCII file but the format is incorrect.
Properly format the ASCII file before converting; refer to Convert,
page for the proper format.
You attempted to convert an ASCII file but one of the particle sizes
read from the input file is less than zero.
Correct the particle sizes in the input file, then convert the file.
Appendix A
539-42822-01 (Rev A) - Nov 2017