Elzone II 5390 Confirm
6133-
Calibration failed: Fit tolerances exceeded.
6134-
Calibration failed: Error in fit too large.
6136-
The automatic calibration could not be performed because the number of ref-
erence materials is greater than the number of peaks. All reference materials
have been removed from the fit.
Cause:
Action:
6151-
This sample file was analyzed on a different instrument. Do you want to pro-
ceed anyway?
Cause:
Action:
6161-
The sample already contains eight tests and cannot be used for an analysis.
Cause:
Action:
6162-
File (file name) cannot be analyzed. It is currently being edited.
Cause:
Action:
A-40
The calibration could not be completed because the line of best fit (ori-
fice tube calibration review screen) is too far from the measured value.
Make sure the correct sample is in the system, the correct SRMs are
selected in the Reference Material list, and the proper analysis condi-
tions are specified for this sample.
You attempted to perform an analysis using a file that has been used for
an analysis with a different instrument.
Click
to continue with the analysis; however, data will be less
Yes
accurate than if the analysis was performed on the same instrument as
the first analysis. Click
No
was calibrated on the current instrument.
You selected
Start/Continue
already contains eight tests.
Delete some of the tests from the file or choose a different file; you can
only perform eight tests with a file.
You selected a file that was still open in an editing window for anal-
ysis.
Finish the changes to the sample file, then save and close the editing
window.
and choose a file containing an orifice that
after a test completed but the current file
539-42822-01 (Rev A) - Nov 2017