Elzone II 5390 Confirm
6118-
The entered Minimum Particle Size (number) cannot be measured using the
calibration selected in Data Setup. A new calibration spanning the desired size
range is recommended.
Cause:
Action A:
Action B:
6119-
The entered Minimum Particle Size (number) requires a log range of (number):
1 using the calibration selected in Data Setup. A new calibration spanning the
desired size range is recommended.
Cause:
Action A:
Action B:
6120-
Unable to read the calibration file (file name).
Cause:
Action:
6121-
Unable to write the calibration file (file name).
Cause:
Action:
A-38
You selected an orifice tube that has not been calibrated to cover the
specified diameter range.
Enter a size range that will accommodate the selected orifice tube.
Recalibrate the orifice tube using reference samples that will cover the
desired range.
The particle size range for the current calibration is not consistent with
the collected data range.
Perform a calibration using the same range as specified in the message.
Then click
Data Setup
on the collected data dialog and update the cal-
ibration.
Use the existing calibration, in which case interpolation will be used to
span the collected data range.
The calibration file selected for loading is not valid or cannot be read.
Confirm the media containing the alleged calibration file has no prob-
lems.
An unexpected error has occurred.
Contact your Micromeritics service representative.