Keithley SourceMeter 2600 Series Reference Manual page 556

Table of Contents

Advertisement

D-28 Status Model
Measurement Event Registers
As shown in
ment Event Status. Commands are summarized in
bits can also be set by using numeric parameter values. For details, see
gramming enable and transition
For example, either of the following commands will set the VOLTAGE_LIMIT
enable bit:
status.measurement.enable = status.measurement.VOLTAGE_LIMIT
status.measurement.enable = 1
When reading a register, a numeric value is returned. The binary equivalent of this
value indicates which bits in the register are set. For details, see
ters" in this appendix. For example, the following command will read the
measurement enable register:
print(status.measurement.enable)
The bits used in the Measurement Event Registers are described as follows:
Measurement Event Current Limit Summary
This summary register set feeds to ILMT bit B1 of the Measurement Event
Register. Used summary bits for this register include B1 (SMUA) and B2 (SMUB).
A set bit indicates that a SMU (A or B) is in current limit (compliance).
Measurement Event Voltage Limit Summary
This summary register set feeds to VLMT bit B0 of the Measurement Event
Register. Used summary bits for this register include B1 (SMUA) and B2 (SMUB).
A set bit indicates that a SMU (A or B) is in voltage limit (compliance).
2600S-901-01 Rev. A / May 2006
Figure
D-5, there are seven register sets associated with Measure-
Bit B0, Voltage Limit (VLMT) — Set bit indicates that the voltage limit was
exceeded. This bit will be updated only when (1) a measurement is taken or
(2) the smuX.source.compliance command is invoked.
Bit B1, Current Limit (ILMT) — Set bit indicates that the current limit was
exceeded. This bit will be updated only when (1) a measurement is taken or
(2) the smuX.source.compliance command is invoked.
Bit B7, Reading Overflow (ROF) — Set bit indicates that an overflow
reading has been detected.
Bit B8, Buffer Available (BAV) — Set bit indicates that there is at least
one reading stored in either or both of the non-volatile reading buffers.
Bit B11, Output Enable (OE) — Set bit indicates that output enable has
been asserted.
Bit B13, Instrument Summary (INST) — Set bit indicates that a bit in the
measurement instrument summary register is set.
Return to
Series 2600 System SourceMeters Reference Manual
Table
registers" in this appendix.
Appendix D topics
D-9. Keep in mind that
"Pro-
"Reading regis-

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents