Keithley SourceMeter 2600 Series Reference Manual page 135

Table of Contents

Advertisement

Series 2600 System SourceMeters Reference Manual
Selecting the Output-off state
Output-off state menu
The OUTPUT configuration menu can be accessed by pressing the CONFIG key
and then the appropriate OUTPUT ON/OFF key. In the configuration menu, select
OFF STATE to display the OUTPUT OFF STATE menu.
With the OUTPUT OFF STATE menu displayed, select the desired output-off
state: HI-Z (high-impedance), NORMAL or ZERO.
Remote programming
Table 3-3
Table 3-3
Commands to select Output-off state
smuX.source.offlimiti = ivalue
smuX.source.offmode = smuX.OUTPUT_NORMAL Selects normal Output-off state.
smuX.source.offmode = smuX.OUTPUT_HIGH_Z
smuX.source.offmode = smuX.OUTPUT_ZERO
* Model 2601/2611: smuX = smua
Model 2602/2612: smuX = smua (Channel A) or smub (Channel B).
NOTE
The OUTPUT OFF STATE menu can also be
accessed by navigating the configuration menu that
is displayed by pressing the CONFIG key.
lists the commands to select the Output-off state.
Command
Return to
DUT Test Connections 3-19
*
Sets 2611/2612 current limit in
normal Output-off state.
Selects high-impedance Output-
off state.
Selects zero Output-off state.
Section 3 topics
2600S-901-01 Rev. A / May 2006
Description

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents