Series 2600 System SourceMeters Reference Manual
Table 16-3
Model 2611/2612 calibration steps
Function
Calibration steps
Voltage Source
1
and Measure
Current Source
1
and Measure
1. Calibrate only the source for the SENSE_CALA sense steps.
2. Steps must be performed in the order shown.
3. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-10. Calibration
polarities must also be set as shown in the procedures.
4. Output must be off before changing to the CALA sense mode.
Calibration commands
Table 16-4
description of these commands, refer to
2
Calibration points
200mV
200mV
2V
2V
20V
200V
100nA
1μA
10μA
100μA
1mA
1mA
10mA
100mA
1A
1.5A
10A
summarizes remote calibration commands. For a more complete
Return to
3
±1e-10, ±180mV
smuX.SENSE_LOCAL
±1e-10, ±180mV
smuX.SENSE_REMOTE
±1e-10, ±1.8V
smuX.SENSE_LOCAL
±1e-10, ±1.8V
smuX.SENSE_CALA
±1e-10, ±18V
smuX.SENSE_LOCAL
±1e-10, ±180V
smuX.SENSE_LOCAL
±1e-10, ±90nA
smuX.SENSE_LOCAL
±1e-10, ±0.9μA
smuX.SENSE_LOCAL
±1e-10, ±9μA
smuX.SENSE_LOCAL
±1e-10, ±90μA
smuX.SENSE_LOCAL
±1e-10, ±0.9mA
smuX.SENSE_LOCAL
±1e-10, ±0.9mA
smuX.SENSE_CALA
±1e-10, ±9mA
smuX.SENSE_LOCAL
±1e-10, ±90mA
smuX.SENSE_LOCAL
±1e-10, ±0.9A
smuX.SENSE_LOCAL
±1e-10, ±1.35A
smuX.SENSE_LOCAL
±1e-10, ±2.4A
smuX.SENSE_LOCAL
Section
12.
Section 16 topics
Calibration 16-7
4
Sense mode
2600S-901-01 Rev. A / May 2006