Keithley SourceMeter 2600 Series Reference Manual page 37

Table of Contents

Advertisement

Series 2600 System SourceMeters Reference Manual
1 CHANNEL A and CHANNEL B (Channel B on 2602/2612 only)
Input/output connections for source, sense, and guard.
2 DIGITAL I/O
Female DB-25 connector. 14 pins for digital input or output, one pin for Output Enable
(2601/2602) or Safety Interlock (2611/2612); +5V and GND pins are also provided.
Use a cable equipped with a male DB-25 connector (Keithley Instruments part number
CA-126-1CA).
3 IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as the Model
7007-1 or Model 7007-2.
4 Cooling exhaust vent
Exhaust vent for internal cooling fan. Keep vent free of obstructions to prevent overheating.
5 Chassis ground
Ground screw for connections to chassis ground.
6 Low noise chassis ground
Ground jack for connecting Output HI or LO to chassis.
7 RS-232
Female DB-9 connector. For RS-232 operation, use a straight-through (not null modem)
DB-9 shielded cable (Keithley Instruments Model 7009-5) for connection to the PC.
8 TSP-Link
Expansion interface that allows a Series 2600 and other TSP-enabled instruments to
trigger and communicate with each other. Use a category 5e or higher LAN crossover
cable (Keithley Instruments part number CA-180-3A).
9 Power module
Contains the AC line receptacle and power line fuse. The instrument can operate on line
voltages of 100V to 240VAC at line frequencies of 50 or 60Hz.
Return to
Section 1 topics
Getting Started
2600S-901-01 Rev. A / May 2006