Keithley SourceMeter 2600 Series Reference Manual page 470

Table of Contents

Advertisement

15-6
Performance Verification
Performing the verification test procedures
Test summary
If the Series 2600 is not within specifications and not under warranty, see the
calibration procedures in
Test considerations
When performing the verification procedures:
2600S-901-01 Rev. A / May 2006
DC voltage output accuracy
DC voltage measurement accuracy
DC current output accuracy
DC current measurement accuracy
Section 16
Be sure to restore factory front panel defaults as outlined above.
Make sure that the test equipment is properly warmed up and connected to
the Series 2600 output terminals (use 4-wire sensing for voltage).
Make sure the Series 2600 is set to the correct source range.
Be sure the Series 2600 output is turned on before making measurements.
Be sure the test equipment is set up for the proper function and range.
Allow the Series 2600 output signal to settle before making a
measurement.
Do not connect test equipment to the Series 2600 through a scanner,
multiplexer, or other switching equipment.
Return to
Series 2600 System SourceMeters Reference Manual
for information on calibrating the unit.