Keithley SourceMeter 2600 Series Reference Manual page 101

Table of Contents

Advertisement

Series 2600 System SourceMeters Reference Manual TSP Programming Fundamentals 2-49
Factory scripts
A factory script is basically the same as a user script, except a factory script is cre-
ated by Keithley Instruments at the factory and is permanently stored
in non-volatile memory. Factory scripts are documented in
Most of the information for
to factory scripts. The differences between a user script and a factory script
include the following:
Running a factory script
Use either of the following commands to run a factory script:
script.factory.scripts.name()
script.factory.scripts.name.run()
where:
Example:
Run the factory script named "sourceMeasureDC":
script.factory.scripts.sourceMeasureDC()
Running a factory script from the Series 2600 front panel controls
1. Press the LOAD key.
2. Select Factory.
3. Select the function to run and press the RUN key.
Modifying a factory script
Retrieving a factory script listing
The script listing for a factory script can be retrieved and modified. However, it
cannot be saved as a factory script. The modified script can be saved as a user
script using the same name or a new name.
"User
A factory script cannot be deleted from non-volatile memory.
The script listing for a factory script can be retrieved and modified, but it will
then be treated as a user script. A user script cannot be saved as a factory
script.
is the name of the factory script.
name
NOTE
An imported factory script can only be loaded back
into the Series 2600 as a user script.
Return to
scripts" (found earlier in this section) also applies
Section 2 topics
2600S-901-01 Rev. A / May 2006
Section
13.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents