Keithley SourceMeter 2600 Series Reference Manual page 478

Table of Contents

Advertisement

15-14 Performance Verification
Figure 15-3
Current verification connections
(2601/2602 3A range; 2611/2612 1.5A range)
Table 15-7
Model 2611/2612 output current accuracy limits
2611/2612
source range
2600S-901-01 Rev. A / May 2006
CAT I
!
CHANNEL B
Model
Model 2611/2612
output
current setting
100nA
90.000nA
1µA
0.90000µA
10µA
9.0000µA
100µA
90.000µA
1mA
0.90000mA
10mA
9.0000mA
100mA
90.000mA
1A
0.90000A
1.5A
1.35000A
10A
9.00000A
Return to
Series 2600 System SourceMeters Reference Manual
W
CHANNEL A
U L
!
CAT I
C
US
!
!
R
Output current limits
(1 year, 18°C–28°C)
89.846 to 90.154nA
0.89893 to 0.90107µA
8.9923 to 9.0077µA
89.913 to 90.087µA
0.89943 to 0.90057mA
8.9913 to 9.0087mA
89.943 to 90.057mA
0.89775 to 0.90225A
1.34519 to 1.35481A
8.958 to 9.042A
Section 15 topics

Hide quick links:

Advertisement

Table of Contents
loading