Keithley SourceMeter 2600 Series Reference Manual page 218

Table of Contents

Advertisement

8-12
Source-Measure Concepts
I -Source operating boundaries
Model 2601/2602 I -Source operating boundaries
Figure 8-5
quadrant of operation is covered. Operation in the other three quadrants is similar.
Figure
Model 2601/2602 SourceMeter can output up to 1.01A at 40V, or 3.03A at 6V.
Note that when sourcing more than 1.01A, voltage is limited to 6V.
Figure
represents the maximum source value possible for the presently selected current
source range. The voltage compliance limit line represents the actual compliance
that is in effect. See
boundaries that represent the operating limits of the SourceMeter for this quad-
rant of operation. The operating point can be anywhere inside (or on) these limit
lines. The limit line boundaries for the other quadrants are similar.
2600S-901-01 Rev. A / May 2006
Figure 8-4
Model 2611/2612 continuous power operating boundaries
shows the operating boundaries for the I-Source. Only the first
8-5A shows the output characteristics for the I-Source. As shown, the
8-5B shows the limit lines for the I-Source. The current source limit line
"Compliance
Return to
Series 2600 System SourceMeters Reference Manual
limit" earlier in this section. These limit lines are
Section 8 topics
£

Hide quick links: