Keithley SourceMeter 2600 Series Reference Manual page 245

Table of Contents

Advertisement

Series 2600 System SourceMeters Reference Manual
Front panel operation
The TSP-Link reset can be performed from the MAIN MENU of any instrument;
this process is summarized in
Table 9-2
Resetting the TSP-Link from the front panel
1) Press the MENU key to access MAIN MENU.
2) Select the COMMUNICATION menu.
3) Select the TSPLINK-CFG menu.
4) Select the RESET to initialize the TSP-Link.
Remote programming
The commands associated with TSP-Link reset are listed in
Table 9-3
TSP-Link reset commands
tsplink.reset()
tsplink.state
An attempted TSP-Link reset will fail if any of the following conditions are true:
Programming example – The following code will reset the TSP-Link and output
its state:
tsplink.reset()
print(tsplink.state)
If the reset is successful,
with all nodes have been established.
Series 2600 MAIN MENU
Command
Initializes the TSP-Link system.
Returns
successful. Returns
Two or more instruments in the system have the same node number.
There are no other instruments connected to the unit performing the reset.
One or more of the units in the system is not powered on.
online
Return to
System Expansion (TSP-Link) 9-5
Table
9-2.
Description
if the most recent TSP-Link reset was
online
offline
will be returned to indicate that communications
Section 9 topics
Table
9-3.
if the reset failed.