Keithley SourceMeter 2600 Series Reference Manual page 486

Table of Contents

Advertisement

16-6
Calibration
Table 16-2
Model 2601/2602 calibration steps
Function
Calibration steps
Voltage Source
1
and Measure
Current Source
1
and Measure
1. Calibrate only the source for the SENSE_CALA sense steps.
2. Steps must be performed in the order shown.
3. 10A range for changing calibration of range only and is not available for normal use.
4. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-10. Calibration
polarities must also be set as shown in the procedures.
5. Output must be off before changing to the CALA sense mode.
2600S-901-01 Rev. A / May 2006
2
Calibration points
100mV
100mV
1V
1V
6V
40V
100nA
1μA
10μA
100μA
1mA
1mA
10mA
100mA
1A
3A
3
10A
Return to
Series 2600 System SourceMeters Reference Manual
4
±1e-10, ±90mV
smuX.SENSE_LOCAL
±1e-10, ±90mV
smuX.SENSE_REMOTE
±1e-10, ±0.9V
smuX.SENSE_LOCAL
±1e-10, ±0.9V
smuX.SENSE_CALA
±1e-10, ±5.4V
smuX.SENSE_LOCAL
±1e-10, ±36V
smuX.SENSE_LOCAL
±1e-10, ±90nA
smuX.SENSE_LOCAL
±1e-10, ±0.9μA
smuX.SENSE_LOCAL
±1e-10, ±9μA
smuX.SENSE_LOCAL
±1e-10, ±90μA
smuX.SENSE_LOCAL
±1e-10, ±0.9mA
smuX.SENSE_LOCAL
±1e-10, ±0.9mA
smuX.SENSE_CALA
±1e-10, ±9mA
smuX.SENSE_LOCAL
±1e-10, ±90mA
smuX.SENSE_LOCAL
±1e-10, ±0.9A
smuX.SENSE_LOCAL
±1e-10, ±2.4A
smuX.SENSE_LOCAL
±1e-10, ±2.4A
smuX.SENSE_LOCAL
Section 16 topics
5
Sense mode