8-14
Source-Measure Concepts
Model 2611/2612 I -Source operating boundaries
Figure 8-6
quadrant of operation is covered. Operation in the other three quadrants is similar.
Figure
Model 2611/2612 SourceMeter can output up to 101mA at 200V, or 1.515A at
20V. Note that when sourcing more than 101mA, voltage is limited to 20V.
Figure
represents the maximum source value possible for the presently selected current
source range. The voltage compliance limit line represents the actual compliance
that is in effect. See
boundaries that represent the operating limits of the SourceMeter for this quad-
rant of operation. The operating point can be anywhere inside (or on) these limit
lines. The limit line boundaries for the other quadrants are similar.
2600S-901-01 Rev. A / May 2006
shows the operating boundaries for the I-Source. Only the first
8-6A shows the output characteristics for the I-Source. As shown, the
8-6B shows the limit lines for the I-Source. The current source limit line
"Compliance
Return to
Series 2600 System SourceMeters Reference Manual
limit" earlier in this section. These limit lines are
Section 8 topics