Keithley SourceMeter 2600 Series Reference Manual page 321

Table of Contents

Advertisement

Series 2600 System SourceMeters Reference Manual
localnode attributes
The attributes in this group are used to set the power line frequency, control (on/off)
prompting and control (hide/show) error messages on the display.
localnode.linefreq
Attribute
Power line frequency.
Usage
frequency = localnode.linefreq
localnode.linefreq = frequency
frequency
• To achieve optimum noise rejection when performing measurements at integer
Remarks
NPLC apertures, the line frequency setting must match the frequency (50Hz or
60Hz) of the AC power line.
• When used in an expanded system (TSP-Link),
the Remote Master node only. Use
ber) to send the command to any node in the system. See
TSP-Link.
• When this attribute is set, the localnode.linefreqauto attribute will be
automatically set to false.
Also see
localnode.linefreqauto
Example
Sets the line frequency to 60Hz:
localnode.linefreq = 60
Set to 50 or 60.
node[N].linefreq
Return to
Section 12 topics
Instrument Control Library 12-47
-- Reads line frequency.
-- Writes line frequency.
localnode.linefreq
(where
is the node num-
N
Section 9
2600S-901-01 Rev. A / May 2006
is sent to
for details on