13-10 Factory Scripts
Example:
2600S-901-01 Rev. A / May 2006
Series 2600 System SourceMeters Reference Manual
SweepVLogMeasureI(smua, 1, 10, 0.001, 5)
This function performs a 5-point logarithmic voltage sweep starting at 1V and
stopping at 10V. Current is measured at every step (point) in the sweep. The
source will be allowed to settle on each step for 1ms before a measurement
is performed.
The following log values and corresponding source levels for the 5-point log
sweep are listed as follows:
MeasurePoint
LogStepSize
1
0.25
2
0.25
3
0.25
4
0.25
5
0.25
Return to
LogStep
0.0
0.25
0.5
0.75
1.0
Section 13 topics
SourceStepLevel
1.0000V
1.7783V
3.1623V
5.6234V
10.000V