Keithley SourceMeter 2600 Series Reference Manual page 404

Table of Contents

Advertisement

12-130 Instrument Control Library
status.system5.*
status.system5.condition
status.system5.enable
status.system5.event
* = condition, enable or event
Attribute
TSP-Link system5 data structure register set.
Usage
Reads condition, enable and event registers:
enablereg = status.system5.condition
enablereg = status.system5.enable
enablereg = status.system5.event
Writes to enable register:
status.system5.enable = enablereg
Set
0
or
1
or
1
status.system5.NODEn
• In an expanded system (TSP-Link), this attribute is used to read or write to the
Remarks
system5 node registers.
• Reading a system5 node register returns a numeric value whose binary equivalent
indicates which register bits are set. The bits of the system5 node register are
identified as follows:
Bit B0 – EXT bit
Bit B1 – Node 57
Bit B2 – Node 58
Bit B3 – Node 59
For example, assume value 9 is returned for the enable register. The binary
equivalent is 0000000000001001. This value indicates that bit 0 (EXT) and bit 3
(Node 59) are set.
• Assigning a value to the
a node bit of the system5 node enable register.
Details
See
"System Summary Event
Also see
status.system.*, status.system2.*, status.system3.*,
Example
Sets the extension bit of the system5 enable register:
status.system5.enable = status.system5.EXT
2600S-901-01 Rev. A / May 2006
to one of the following values:
enablereg
status.system5.EXTENSION_BIT
status.system5.EXT
Bit B4 – Node 60
Bit B5 – Node 61
Bit B6 – Node 62
Bit B7 – Node 63
status.system5.enable
Return to
Series 2600 System SourceMeters Reference Manual
Clears all bits.
Sets EXT bit (B0).
Sets EXT bit (B0).
Sets a node bit (Bn);
Bit B8 – Node 64
Bits B9 through B15 – Not used
Registers" in
Appendix
Section 12 topics
= 57 to 64.
n
attribute sets the extension bit or
D.
status.system4.*