Keithley SourceMeter 2600 Series Reference Manual page 422

Table of Contents

Advertisement

13-12 Factory Scripts
TSP Project Name:
TSP Test Script Name:
Firmware Version:
Function:
SweepVListMeasureI(smu, vlist, stime, points)
Description: Performs a voltage list sweep with current measured at every step (point):
Parameters:
Data: Current measurements, voltage source values and timestamps are stored in
Example:
2600S-901-01 Rev. A / May 2006
Series 2600 System SourceMeters Reference Manual
KIFactoryGeneral
KIGeneral
1.0.2 and later
• Sets the
to output the first
smu
settle for
seconds and then performs a current measurement.
stime
• Sets the
to output the next
smu
settle for
seconds and then performs a current measurement.
stime
• Repeats the above sequence until the current is measured for the last
volts value. The last point in the list to be measured is
smu, vlist, stime, points
SourceMeter Channel (A or B). Defaults to smua if all parameters
smu:
are omitted when function is called.
Arbitrary list of voltage source values:
vlist:
vlist = {value1, value2, ....valueN}
Settling time in seconds. Occurs after sourcing a value and before
stime:
performing a measurement.
Number of sweep measure points (must be
points:
.
smuX.nvbuffer1
myvlist = {-0.1, 0.1, -1, 1, -6, 6, -40, 40, 0, 0}
SweepVListMeasureI(smua, myvlist, 500E-6, 10)
This function performs a 10-point voltage list sweep starting at the first point
in
. Current is measured at every step (point) in the sweep. The
myvlist
source will be allowed to settle on each value for 500µs before a
measurement is performed.
Return to
volts value, allows the source to
vlist
volts value, allows the source to
vlist
Section 13 topics
.
points
≥ 2).