Series 2600 System SourceMeters Reference Manual
TSP Project Name:
TSP Test Script Name:
Firmware Version:
Function:
SweepILinMeasureV(smu, starti, stopi, stime, points)
Description: Performs a linear current sweep with voltage measured at every step (point):
Parameters:
Data: Voltage measurements, current source values and timestamps are stored in
Example:
KIFactoryGeneral
KIGeneral
1.0.2 and later
• Sets the
to output
smu
seconds and then performs a voltage measurement.
stime
• Sets the
to output the next amps step, allows the source to settle for
smu
seconds and then performs a voltage measurement.
stime
• Repeats the above sequence until the voltage is measured on the
amps step.
The linear step size is automatically calculated as follows:
step = (
stopi
starti
starti
stime
smu, starti, stopi, stime, points
SourceMeter Channel (A or B). Defaults to smua if all parameters
smu:
are omitted when function is called.
Sweep start current in amps.
starti:
Sweep stop current in amps.
stopi:
Settling time in seconds. Occurs after stepping the source and
stime:
before performing a measurement.
Number of sweep points (must be
points:
.
smuX.nvbuffer1
SweepILinMeasureV(smua, -1E-3, 1E-3, 0, 100)
This function performs a 100-point linear current sweep starting at -1mA and
stopping at +1mA. Voltage is measured at every step (point) in the sweep.
Since
is set for 0s, voltage will be measured as fast as possible after
stime
each current step.
Return to
amps, allows the source to settle for
starti
) / (
– 1)
points
stopi
stime
stime
≥ 2).
Section 13 topics
2600S-901-01 Rev. A / May 2006
Factory Scripts 13-5
stopi