12-92 Instrument Control Library
smuX.source.outputenableaction
Attribute
Output enable or interlock action for the source.
Usage
action = smuX.source.outputenableaction
smuX.source.outputenableaction = action
Set
or
0
or
1
• This attribute controls the SMU action taken when the output enable or interlock line
Remarks
is asserted/de-asserted. The default setting is
• This attribute is always set to OE_OUTPUT_OFF for the Model 2611/2612 and is
read-only.
• When set to
interlock line goes low (de-asserted).
• When set to
asserted, the SMU will turn its output off as if the
smuX.OUTPUT_OFF
command had been received.
• The SMU will not automatically turn its output on when the output enable or interlock
line returns to the high state.
• If the output enable or interlock line is not asserted when this attribute is set to
smuX.OE_OUTPUT_OFF
• Detection of the output enable or interlock line going low will not abort any running
scripts. This may cause execution errors.
Details
See
"Output Enable (Models
Also see
smuX.source.offmode,
Example
Reconfigures the SMU to turn the output off if the output enable or interlock line goes
low (de-asserted):
smua.source.outputenableaction = smua.OE_OUTPUT_OFF
2600S-901-01 Rev. A / May 2006
to one of the following values:
action
smuX.OE_NONE
smuX.OE_OUTPUT_OFF
smuX.OE_NONE
smuX.OE_OUTPUT_OFF
and the output is on, the output will turn off immediately.
smuX.source.output
Return to
Series 2600 System SourceMeters Reference Manual
X = SMU channel (a or b)
No action.
Turns the source output off.
smuX.OE_NONE
, the SMU will take no action when the output enable or
and the output enable or interlock line is de-
2601/2602)" in
Section
Section 12 topics
-- Reads enable action.
-- Writes enable action.
.
smuX.source.output =
10.