13-20 Factory Scripts
Waveform:
Return
Values:
Output
Data:
Example:
See Also: InitiatePulseTest,
2600S-901-01 Rev. A / May 2006
Series 2600 System SourceMeters Reference Manual
step = (stop - start)/(points -1)
start
bias
A Boolean flag. This flag will be true when the pulse was
f:
successfully configured, false when errors were encountered.
A string message. If the f flag is false, msg will contain an error
msg:
message. Otherwise, msg will contain a string indicating successful
configuration.
Pulsed voltage measurements are stored in the reading buffer specified by
the buffer input argument.
ConfigPulseIMeasureVSweepLin(smua, 0, 0.01, 0.05, 1, 1e-3,
0.1, 20, smua.nvbuffer2, 3)
Setup a pulsed sweep that will use SourceMeter channel A. The pulsed
sweep will start at 10mA, end at 50mA, and return to a 0mA bias level
between pulses. Each pulsed-step will be on for 1msec and then at the bias
level for 100msec. The voltage limit will be 1V during the entire pulsed-
sweep. The pulse train will be comprised of 20 pulsed-steps and the pulse
train will be assigned a tag index of 4.
Return to
step
step
ton
toff
InitiatePulseTestDual
Section 13 topics
stop