Series 2600 System SourceMeters Reference Manual
Figure 15-1
Connections for voltage verification
Table 15-2
Model 2601/2602 output voltage accuracy limits
Model 2601/2602
source range
Table 15-3
Model 2611/2612 output voltage accuracy limits
Model 2611/2612
source range
CAT I
!
CHANNEL B
Model 2601/2602
output voltage setting
100mV
90.000mV
1V
0.90000V
6V
5.4000V
40V
36.000V
Model 2611/2612
output voltage setting
200mV
180.000mV
2V
1.80000V
20V
18.000V
200V
180.000V
Return to
CHANNEL A
U L
!
CAT I
C
US
!
!
R
Output voltage limits
Output voltage limits
(1 year, 18°C–28°C)
Section 15 topics
2600S-901-01 Rev. A / May 2006
Performance Verification 15-9
(1 year, 18°C–28°C)
89.732 to 90.268mV
0.89942 to 0.90058V
5.39712 to 5.40288V
35.9808 to 36.0192V
179.589 to 180.411mV
1.79904 to 1.80096V
17.9914 to 18.0086V
179.914 to 180.086V