12-72 Instrument Control Library
smuX.measure.filter.count
Attribute
Number of measured readings to yield one filtered measurement.
Usage
count = smuX.measure.filter.count
smuX.measure.filter.count = count
count
• This attribute is the number of measurements that will be performed to yield one
Remarks
filtered measurement.
• The
Details
See "Filters" in
Also see
smuX.measure.filter.enable,
Sets filter count for SMU A:
Example
smua.measure.filter.count = 10
smuX.measure.filter.enable
Attribute
Enables/disables filtered measurements.
Usage
filter = smuX.measure.filter.enable
smuX.measure.filter.enable = filter
Set
or
0
or
1
• This attribute enables or disables the filter.
Remarks
• The
Details
See "Filters" in
Also see
smuX.measure.filter.count,
Example
Enable the filter for SMU A:
smua.measure.filter.enable = smua.FILTER_ON
smuX.measure.filter.type
Attribute
Type of filter for measurements.
Usage
type = smuX.measure.filter.type
smuX.measure.filter.type = type
Set
or
0
or
1
2600S-901-01 Rev. A / May 2006
Set filter count from 1 to 100.
function sets the filter count to 1.
reset
Section
6.
to one of the following values:
filter
smuX.FILTER_OFF
smuX.FILTER_ON
function disables the filter.
reset
Section
6.
smuX.measure.filter.type
to one of the following values:
type
smuX.FILTER_MOVING_AVG
smuX.FILTER_REPEAT_AVG
Return to
Series 2600 System SourceMeters Reference Manual
X = SMU channel (a or b)
-- Reads filter count.
-- Writes filter count.
smuX.measure.filter.type
X = SMU channel (a or b)
-- Reads on/off state of the filter.
-- Writes on/off state of the filter.
Disables the filter.
Enables the filter.
X = SMU channel (a or b)
-- Reads filter type.
-- Writes filter type.
Selects the moving average filter.
Selects the repeat filter.