4-12
Basic Operation
Remote triggering
Remote trigger commands
Trigger commands are listed in
The trigger event detector remembers if an event has been detected since the last
trigger.wait call. The trigger.clear() command clears the trigger's
event detector and discards the previous history of command interface trigger
events.
The trigger.wait function will wait timeout seconds for a GPIB GET
command (see
interface if that is the active command interface, or a *TRG message on the
command interface for all other interfaces. If one or more of these trigger events
were previously detected, this function will return immediately. After waiting for a
trigger with this function, the event detector will be automatically reset and re-
armed regardless of the number of events detected.
Table 4-6
Trigger commands
trigger.clear()
triggered = trigger.wait(timeout)
Remote trigger example
The example below clears triggers and then enables a 30 second timeout to wait
for a GET or *TRG trigger.
smua.reset()
triggered = trigger.wait(30)
smua.source.output =smua.OUTPUT_ON
*TRG
reading = smua.measure.v()
2600S-901-01 Rev. A / May 2006
Table
Section
11) or a *TRG message (see
Command
Return to
Section 4 topics
Series 2600 System SourceMeters Reference Manual
4-6. See Section 12 for more details.
Appendix
Description
Clear pending triggers.
Wait timeout seconds for GET or *TRG trigger.
--Restore 2600 defaults.
--Wait 30s for GET or *TRG.
--Turn on output.
--Trigger reading.
--Get voltage reading.
C) on the GPIB