Keithley SourceMeter 2600 Series Reference Manual page 266

Table of Contents

Advertisement

11-6
Communications Interfaces
Front panel primary address
To set or check the primary address:
1.
2.
3.
4.
5.
6.
Remote primary address
Use the following command to set the primary address by remote:
gpib.address = address
For example, the following command sets the address to 20:
gpib.address = 20
Note that changing the GPIB address takes effect when the command is
processed. Any response messages generated after processing this command
will be sent with the new settings. If command messages are being queued (sent
before this command has executed), the new settings may take effect in the
middle of a subsequent command message, so care should be exercised when
setting this attribute from the remote interface.
Terminator
When receiving data over the GPIB, the Series 2600 will terminate on any line
feed character or any data byte with EOI asserted (line feed with EOI asserted is
also valid). When sending data, it will append a line feed character to all outgoing
messages. The EOI line will be asserted with the terminating line feed character.
2600S-901-01 Rev. A / May 2006
Press the MENU key.
Select COMMUNICATIONS, then press ENTER or the Rotary Knob.
Select INTERFACE_CFG, then press ENTER or the Rotary Knob.
Select GPIB, then press ENTER or the Rotary Knob.
Set the primary address to the desired value, then press ENTER or the
Rotary Knob.
Press EXIT to back out of the menu structure.