Keithley SourceMeter 2600 Series Reference Manual page 150

Table of Contents

Advertisement

4-14
Basic Operation
4.
Step 3: Select measurement function and range.
Select measurement function and range as follows:
1.
2.
Step 4: Turn output on.
Turn the output on by pressing the ON/OFF OUTPUT key. The OUTPUT indicator
light will turn on.
Step 5: Observe readings on the display.
Observe the readings on the display. Press TRIG if necessary to trigger the unit to
begin taking readings. For the single-channel display mode, the readings will
appear on the top line, while source and limit values are on the bottom line.
Step 6: Turn output off.
When finished, turn the output off by pressing the ON/OFF OUTPUT key. The
OUTPUT indicator light will turn off.
Remote source-measure procedure
Basic source-measurement procedures can also be performed via remote by
sending appropriate commands in the right sequence. The following table
summarizes the basic commands and give a simple programming example.
2600S-901-01 Rev. A / May 2006
Enter the desired limit value, then press ENTER or the Rotary Knob to
complete editing.
Put the Model 2602/2612 in the single-channel display mode, then select
the desired measurement function by pressing MEAS or MODE.
Select the desired measurement range with the RANGE keys, or enable
AUTO RANGE, keeping the following points in mind:
When measuring the source (i.e., Source V Measure V), you cannot
select the measurement range using the RANGE keys. The selected
source range determines the measurement range.
When not measuring the source (i.e., Source V Measure I),
measurement range selection can be done manually or automatically.
When using manual ranging, use the lowest possible range for best
accuracy. In auto range, the SourceMeter automatically goes to the
most sensitive range to make the measurement.
Return to
Series 2600 System SourceMeters Reference Manual
Section 4 topics

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents