Series 2600 System SourceMeters Reference Manual
Table 4-1
Source-measure capabilities
Range
100mV
1V
6V
40V
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
3A
Max Power = 40.4W per channel
Compliance limit
When sourcing voltage, the SourceMeter can be set to limit current. Conversely,
when sourcing current, the SourceMeter can be set to limit voltage. The
SourceMeter output will not exceed the compliance limit. The maximum
compliance limit is the same as the maximum values listed in
the compliance value will take the same sign as the source value, and the
maximum compliance limits are based on source range. See
Section 8
being exceeded is the VLIMIT when operating as an ISOURCE. To avoid
excessive (and potentially destructive) currents from flowing, the VLIMIT will
source or sink up to 102mA for ISOURCE ranges on or below 100mA. For the
ranges 1A and above, the maximum current allowed is the current source setting.
Model 2601/2602
Source
Measure
±101mV
±102mV
±1.01V
±1.02V
±6.06V
±6.12V
±40.4V
±40.8V
±101nA
±102nA
±1.01µA
±1.02µA
±10.1µA
±10.2µA
±101µA
±102µA
±1.01mA
±1.02mA
±10.1mA
±10.2mA
±101mA
±102mA
±1.01A
±1.02A
±3.03A
±3.06A
for more information. The only exception to the compliance limit not
Return to
Model 2611/2612
Range
Source
200mV
±202mV
2V
±2.02V
20V
±20.2V
1
200V
±202V
100nA
±101nA
1µA
±1.01µA
10µA
±10.1µA
100µA
±101µA
1mA
±1.01mA
10mA
±10.1mA
100mA
±101mA
1A
±1.01A
1.5A
±1.515A
2
10A
±10.1A
Max Power = 30.603W per channel
1. 200V source range available only when
interlock is enabled. See
2. 10A range available only in pulse mode.
Section 4 topics
2600S-901-01 Rev. A / May 2006
Basic Operation
4-3
Measure
±204mV
±2.04V
±20.4V
±204V
±102nA
±1.02µA
±10.2µA
±102µA
±1.02mA
±10.2mA
±102mA
±1.02A
±1.53A
±10.2A
Section
10.
Table
4-2. Note that
"Compliance
limit" in