Keithley SourceMeter 2600 Series Reference Manual page 174

Table of Contents

Advertisement

5-12
Sweep Operation
Linear staircase sweep example
1.
2.
3.
4.
5.
2600S-901-01 Rev. A / May 2006
Configure source functions.
Examples – The following commands restore defaults and set the
compliance to 1V:
smua.reset()
smua.source.limitv = 1
Configure the sweep.
Example – The following parameters configure a linear staircase current
sweep from 1mA to 10mA with 10 points and a 0.1 second settling time:
SweepILinMeasureV(smua, 1e-3,
10e-3, 0.1, 10)
Run the sweep.
smua.source.output =
smua.OUTPUT_ON
waitcomplete()
Turn off the output.
When the sweep is done, turn the source output off with this
command:
smua.source.output = smua.OUTPUT_OFF
Request readings.
Request readings from Buffer 1 as follows:
printbuffer(1, 10, smua.nvbuffer1.readings)
Return to
Series 2600 System SourceMeters Reference Manual
--Restore Series 2600 defaults.
--Set compliance to 1V.
--Linear staircase sweep,
channel A, 1mA to 10mA, 0.1
second delay, 10 points.
--Turn on output.
--Wait for sweep to complete.