System behavior
Using the reset() command
While most TSP-Link operations target a single node in the system, the reset()
command affects the system as a whole. The reset() command, by definition,
resets all nodes to their default settings:
reset()
node[N]
The other nodes are not affected. Examples:
node[1].reset()
localnode.reset()
node[4].reset()
Abort
An abort will terminate an executing script and return all nodes to local operation
(REM annunciators turn off), dissolving the Master/Slave relationships between
nodes. An abort is invoked by either issuing an abort command to the Master or
pressing the EXIT key on any node in the system.
An abort can also be performed by pressing the OUTPUT ON/OFF key on any
node. The results are the same as above, with the addition that all SMU outputs in
the system are turned off.
-- Resets all nodes in a TSP-Link system.
and localnode can be used with reset to reset only one of the nodes.
-- Resets Node 1 only.
-- Resets Node 1 only.
-- Resets Node 4 only.