16-10 Calibration
Step 2. Voltage Calibration
1.
2.
Figure 16-1
Connections for voltage calibration
2600S-901-01 Rev. A / May 2006
Connect the Series 2600 to the digital multimeter using the 4-wire
connections shown in
function.
Send the following commands in order to initialize voltage calibration:
smua.cal.unlock("KI002602")
smua.reset()
smua.source.func = smua.OUTPUT_DCVOLTS
(Substitute the actual calibration password in the smua.cal.unlock
command. The default Model 2601 password is "KI002601"; the default
Model 2602 password is "KI002602". The default Model 2611 password is
"KI002611"; the default Model 2612 password is "KI002612".)
CAT I
!
CHANNEL B
Return to
Series 2600 System SourceMeters Reference Manual
Figure
16-1, and select the multimeter DC volts
CHANNEL A
U L
!
CAT I
C
US
!
!
R