Keithley SourceMeter 2600 Series Reference Manual page 508

Table of Contents

Advertisement

2601
2602
SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the Models 2601 and 2602.
Specifications are the standards against which the Models 2601 and 2602 are tested. Upon leaving the
factory the 2601 and 2602 meet these specifications. Supplemental and typical values are non-
warranted, apply at 23°C, and are provided solely as useful information.
The source and measurement accuracies are specified at the SourceMeter CHANNEL A (2601 and
2602) or SourceMeter CHANNEL B (2602) terminals under the following conditions:
1. 23°C ± 5°C, <70% relative humidity.
2. After 2 hour warm-up.
3. Speed normal (1 NPLC).
4. A/D auto-zero enabled.
5. Remote sense operation or properly zeroed local operation.
6. Calibration period = 1 year.
SOURCE SPECIFICATIONS
VOLTAGE PROGRAMMING ACCURACY
PROGRAMMING
RANGE
RESOLUTION
100.000 mV
5 µV
1.00000 V
50 µV
6.00000 V
50 µV
40.0000 V
500 µV
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15 × accuracy specification)/°C.
MAXIMUM OUTPUT POWER AND SOURCE/SINK LIMITS
@ ±1.0A, ±6.06V @ ±3.0A, four quadrant source or sink operation.
VOLTAGE REGULATION: Line: 0.01% of range. Load: ±(0.01% of range + 100µV).
NOISE 10Hz–20MHz (peak-peak): 25mV typical into a resistive load.
CURRENT LIMIT/COMPLIANCE
3
: Bipolar current limit (compliance) set with single value. Minimum
value is 10nA. Accuracy same as current source.
OVERSHOOT: <±(0.1% + 10mV) typical (step size = 10% to 90% of range, resistive load, maximum
current limit/compliance).
GUARD OFFSET VOLTAGE: <10mV typical (Iout ≤ 100mA).
CURRENT PROGRAMMING ACCURACY
PROGRAMMING
RANGE
RESOLUTION
100.000 nA
1 pA
1.00000 µA
10 pA
10.0000 µA
100 pA
100.000 µA
1 nA
1.00000 mA
10 nA
10.0000 mA
100 nA
100.000 mA
1 µA
1.00000 A
2
10 µA
3.00000 A
2
10 µA
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15 × accuracy specification)/°C.
MAXIMUM OUTPUT POWER AND SOURCE/SINK LIMITS
@ ±40.0V, ±3.03A @ ±6.0V, four quadrant source or sink operation.
CURRENT REGULATION: Line: 0.01% of range. Load: ±(0.01% of range + 100pA).
VOLTAGE LIMIT/COMPLIANCE
4
: Bipolar voltage limit (compliance) set with a single value.
Minimum value is 10mV . Accuracy same as voltage source.
OVERSHOOT: <0.1% typical (step size = 10% to 90% of range, resistive load; see CURRENT SOURCE
OUTPUT SETTLING TIME for additional test conditions).
A-2
System SourceMeter
Multi-Channel I-V Test Solutions
1
ACCURACY (1 Year)
TYPICAL NOISE
23°C ±5°C
(peak-peak)
±(% rdg. + volts)
0.1Hz–10Hz
0.02% + 250 µV
20 µV
0.02% + 400 µV
50 µV
0.02% + 1.8 mV
100 µV
0.02% + 12 mV
500 µV
2
: 40.4W per channel maximum. ±40.4V
ACCURACY (1 Year)
TYPICAL NOISE
23°C ±5°C
(peak-peak)
±(% rdg. + amps)
0.1Hz–10Hz
0.06% + 100 pA
5 pA
0.03% + 600 pA
25 pA
0.03% +
2 nA
50 pA
0.03% + 30 nA
3 nA
0.03% + 200 nA
5 nA
0.03% +
3 µA
200 nA
0.03% + 20 µA
500 nA
0.05% + 900 µA
60 µA
0.06% + 1.5 mA
150 µA
2
: 40.4W per channel maximum. ±1.01A
®
ADDITIONAL SOURCE SPECIFICATIONS
TRANSIENT RESPONSE TIME: <70µs for the output to recover to 0.1% for a 10% to 90% step
change in load.
VOLTAGE SOURCE OUTPUT SETTLING TIME: Time required to reach 0.1% of final value, when
changing from 10% to 90% of range, after source level command is processed on a fixed range.
100mV, 1V Ranges: <50µs typical.
6V Range: <100µs typical.
40V Range: <150µs typical.
5
CURRENT SOURCE OUTPUT SETTLING TIME: Time required to reach 0.1% of final value, when
changing from 10% to 90% of range, after source level command is processed on a fixed range. Values
below for Iout · Rload = 2V unless noted.
3A–10mA Ranges: <80µs typical (current less than 2.5A, Rload >1.5Ω).
1mA Range: <100µs typical.
100µA Range: <150µs typical.
10µA Range: <500µs typical.
1µA Range: <2.5ms typical.
100nA Range: <25ms typical.
DC FLOATING VOLTAGE: Output can be floated up to ±250VDC from chassis ground.
REMOTE SENSE OPERATING RANGE
1
:
Maximum voltage between HI and SENSE HI = 3V.
Maximum voltage between LO and SENSE LO = 3V.
VOLTAGE OUTPUT HEADROOM:
40V Range: Max. output voltage = 42V – total voltage drop across source leads (maximum 1Ω per
source lead).
6V Range: Max. output voltage = 8V – total voltage drop across source leads.
OVER TEMPERATURE PROTECTION: Internally sensed temperature overload puts unit in
standby mode.
VOLTAGE SOURCE RANGE CHANGE OVERSHOOT: Overshoot into a 100kΩ load, 20MHz BW ,
300mV typical.
CURRENT SOURCE RANGE CHANGE OVERSHOOT: <5% + 300mV/Rload of larger range typical.
(See CURRENT SOURCE OUTPUT SETTLING TIME for additional test conditions.)
NOTES
1. Add 50µV to source accuracy specifications per volt of HI lead drop.
2. Full power source operation regardless of load to 30°C ambient. Above 30°C and/or power sink operation, refer to
Section 8 – Operating Boundaries in the Series 2600 Reference Manual for additional power derating information.
3. For sink mode operation (quadrants II and IV), add 12% of limit range and ±0.02% of limit setting to corresponding
current limit accuracy specifications. For 1A range add an additional 40mA of uncertainty.
4. For sink mode operation (quadrants II and IV), add 10% of compliance range and ±0.02% of limit setting to corre-
sponding voltage source specification. For 100mV range add an additional 60mV of uncertainty.
5. Add 150µs when measuring on the 1A range.