16-4
Calibration
Recommended calibration equipment
Table 16-1
can use alternate equipment as long as that equipment has specifications equal to
or greater than those listed in the table. When possible, test equipment
specifications should be at least four times better than corresponding Series 2600
specifications.
Table 16-1
Recommended calibration equipment
Description
Digital Multimeter
0.5Ω, 250W, 0.1%
Precision Resistor
50Ω Resistors (2)
1. 90-day specifications show full-range accuracy of recommended model used for specified
calibration point.
2. Resistor used to calibrate Model 2601/2602 3A and 10A ranges and Model 2611/2612 1.5A and
10A ranges should be characterized to uncertainty shown using resistance function of a digital
multimeter before use.
3. Used for contact check calibration. Characterize resistors using ohms function of digital
multimeter before use.
2600S-901-01 Rev. A / May 2006
lists the recommended equipment for the calibration procedures. You
Manufacturer/
Model
Agilent 3458A
Isotek
RUG-Z-R500-0.1-TK3
Any suitable.
Return to
Section 16 topics
Series 2600 System SourceMeters Reference Manual
Accuracy
DC Voltage
(2601/2602)
DC Voltage
(2611/2612)
1
DC current
Resistance
3
1
90mV:
±8ppm
0.9V:
±5ppm
5.4V:
±4ppm
36V:
±6ppm
1
190mV:
±5ppm
1.8V:
±4ppm
18V:
±6ppm
180V:
±6ppm
90nA:
±430ppm
0.9µA:
±45ppm
9µA:
±25ppm
90µA:
±23ppm
0.9mA:
±20ppm
9mA:
±20ppm
90mA:
±35ppm
0.9A:
±110ppm
2
0.5Ω:
±125ppm