Series 2600 System SourceMeters Reference Manual
Table 8-1
Maximum compliance limits
Source
100mV
100mA
Compliance principles
Compliance acts as a clamp. If the output reaches the compliance value, the
SourceMeter will attempt to prevent the output from exceeding that value. This
action implies that the source will switch from a V-source to an I-source (or from
an I-source to a V-source) when in compliance.
As an example, assume the following:
SourceMeter: V
DUT resistance: 10Ω
With a source voltage of 10V and a DUT resistance of 10Ω, the current through
the DUT should be: 10V/10Ω = 1A. However, because the compliance is set to
10mA, the current will not exceed that value, and the voltage across the
resistance is limited to 100mV. In effect, the 10V voltage source is transformed
into a 10mA current source with a 100mV compliance value.
Model 2601/2602
Maximum
compliance
range
value
3A
1V
3A
6V
3A
40V
1A
100nA
40V
1µA
40V
10µA
40V
100µA
40V
1mA
40V
10mA
40V
40V
1A
40V
3A
6V
= 10V; I
SRC
Return to
Model 2611/2612
Source
range
200mV
2V
20V
200V
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
1.5A
= 10mA
CMPL
Section 8 topics
Source-Measure Concepts
Maximum
compliance
value
1.5A
1.5A
1.5A
100mA
200V
200V
200V
200V
200V
200V
200V
20V
20V
2600S-901-01 Rev. A / May 2006
8-3