Series 2600 System SourceMeters Reference Manual
Sweep Operation
5-7
Figure 5-4
Pulse sweep example
Custom (list) sweeps
This sweep type lets you configure a customized sweep with arbitrary steps.
Programmable parameters include the source function, channel, list of sweep
values, delay (settling time), and the number of points.
When this sweep is started, the output level goes to the first point in the sweep.
The sweep will continue through the steps in the order they were programmed
and stop after the last step. The time duration before the measurement at each
step is determined by the delay interval (settling time).
Figure 5-5
shows an example of a custom sweep with six measurement points.
When the sweep starts, the current or voltage goes to the first point in the sweep.
The unit cycles through the sweep points in the programmed order.
Figure 5-5
Custom sweep example
Return to
Section 5 topics
2600S-901-01 Rev. A / May 2006