Keithley SourceMeter 2600 Series Reference Manual page 365

Table of Contents

Advertisement

Series 2600 System SourceMeters Reference Manual
Details
See
"Output-off
Also see
smuX.source.output
Sets output-off mode for SMU A:
Example
smua.source.offmode = smua.OUTPUT_HIGH_Z
smuX.source.output
Attribute
Source output control (on/off).
Usage
state = smuX.source.output
smuX.source.output = state
Set
state
or
0
or
1
Remarks
Reading this attribute gives the output state of the source. Setting this attribute will turn
the output of the source on or off. The default for the source is off. When the output is
turned on, the SMU will source either voltage or current as dictated by the
smuX.source.func
Details
See
Section 4
Also see
smuX.source.func,
Example
Turns SMU A source output on:
smua.source.output = smua.OUTPUT_ON
states" in
Section
X = SMU channel (a or b)
to one of the following values:
smuX.OUTPUT_OFF
smuX.OUTPUT_ON
setting.
(basic operation).
smuX.source.offmode
Return to
Instrument Control Library 12-91
3.
-- Reads output state.
-- Writes output state.
Turns the source output off.
Turns the source output on.
Section 12 topics
2600S-901-01 Rev. A / May 2006

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents