16-18 Calibration
7.
8.
Figure 16-3
Connections for current calibration
(2601/2602 3A and 10A ranges; 2611/2612 1.5A and 10A ranges)
2600S-901-01 Rev. A / May 2006
Repeat steps a through v for the 3A and 10A ranges (Model 2601/2602) or
1.5A and 10A ranges (Model 2611/2612). Compute the current reading
from the DMM voltage reading and characterized 0.5Ω resistance value: I =
V/R.
Select auto polarity mode:
smua.cal.polarity = smua.CAL_AUTO