Keithley SourceMeter 2600 Series Reference Manual page 475

Table of Contents

Advertisement

Series 2600 System SourceMeters Reference Manual
Table 15-4
Model 2601/2602 voltage measurement accuracy limits
Model 2601/2602 source
and measure range
1. Measure range coupled to source range when simultaneously sourcing and measuring voltage.
2. As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
Table 15-5
Model 2611/2612 voltage measurement accuracy limits
Model 2611/2612 source
and measure range
1. Measure range coupled to source range when simultaneously sourcing and measuring voltage.
2. As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
1
Source voltage
100mV
1V
6V
40V
1
Source voltage
200mV
2V
20V
200V
Return to
Section 15 topics
Performance Verification 15-11
Model 2601/2602 voltage
2
(1 year, 18°C–28°C)
90.000mV
89.8365 to 90.1635mV
0.90000V
0.899665 to 0.900335V
5.4000V
5.39819 to 5.40181V
36.000V
35.9866 to 36.0134V
Model 2611/2612 voltage
2
(1 year, 18°C–28°C)
180.000mV
179.748 to 180.252mV
1.80000V
1.79938 to 1.80062V
18.0000V
17.9923 to 18,0077V
180.000V
179.923 to 180.077V
2600S-901-01 Rev. A / May 2006