Preparation

Assembly of Components

Assembling the expiratory valve

Installing the expiratory valve

Preparation
WARNING!
Do not place containers of liquids (such as infusion
bottles) on top of or above the Savina ventilator. Liquids
getting into the ventilator can cause equipment
malfunction with the risk of patient injury.
Assembly of Components
WARNING!
Always use components that have been properly cleaned
and disinfected
Assembling the expiratory valve
1 Install membrane on expiratory valve body
2 Push rubber grommet onto expiratory valve port
Installing the expiratory valve
3 Turn knurled knob counterclockwise to stop.
4 Insert expiratory valve.
3 Hand-tighten knurled knob clockwise.
5 Install jar of water trap.