Required Adjustment after Replacing Assembly

After replacing the following assembly, the adjustment items described in Table 3-1 must
be required. The adjustment must be performed Agilent's qualified service personnel. If
you need the adjustment for your E4980A, it should be sent to the nearest Agilent
Technologies service office.
Table 3-1
Required Adjustment Item after Replacing Assembly
Replaced Assembly
PPMC PCA
A1 Board
A2 Board
A3 Board

Writing ID

This item writes the serial number, option information into the A1 board of the E4980A.

Write Null Phase Track Default Data

The purpose of this procedure is to write the null phase track default data into A2 board

Null DC Offset Adjustment

The purpose of this procedure is to adjust the null dc offset of the A2 board.

Null ADC DC Offset Adjustment

The purpose of this procedure is to adjust the null adc dc offset of the A2 board.

ADC Linearity Adjustment

The purpose of this procedure is to ajust ADC non-linearity of the A2 board.
Chapter 3
Required Adjustment after Replacing Assembly
Adjustment Item
Adjustment
23